Water Quality Comparison Report for Security Force Makkah

Gemini Link

This report compares the Pre-RO (Softened) and Post-RO (DI/RO) water analysis results against the **Optimal Chemical Water Qualities for Instrument Reprocessing** standard from the GETINGE document.


📞 Contact & Report Summary

Phone: +966-56-470-0751
📧 Email: info@oceangates.sa
🌐 Website: www.ums.sa

Water Treatment System Performance Report - URGENT

Service Call Reason: Received a call from NMG complaining about the performance of the water treatment system.

Visit Date: October 15, 2025
Lab Sample Date: October 18, 2025


Conclusion: The complaint is validated. The R.O. system fails to produce Critical Water compliant with AAMI ST108 for CSSD sterilizer feed.

1. Pre-RO (Softened) Water Quality Comparison

Comparison of Pre-RO water (Sample W.CH.25.214) against the **Softened** water standard.

Parameter Unit of Measure Standard (Softened) Lab Result (Pre-RO) Result
pH value - 6.5-8.5 6.68 PASS
Hardness mg/L (ppm) < 10 2 PASS
Conductivity (@ 25°C) µS/cm < 500 433 PASS
Chloride (Cl) mg/L (ppm) < 60 115 FAIL
Iron mg/L (ppm) < 0.3 0.06 PASS
Copper mg/L (ppm) < 0.1 < 0.001 PASS
Silica (SiO₂) mg/L (ppm) ≤ 2 0 PASS
Phosphate (P₂O₅) mg/L (ppm) < 10 0.03 PASS

Summary for Pre-RO (Softened) Water:

The **Pre-RO (Softened)** water **FAILS** the standard due to a high **Chloride** concentration (115 mg/L), which exceeds the standard limit (< 60 mg/L). All other tested parameters are within the acceptable limits.


2. Post-RO (DI/RO) Water Quality Comparison

Comparison of Post-RO water (Sample W.CH.25.213) against the **DI/RO** water standard.

Parameter Unit of Measure Standard (DI/RO) Lab Result (Post-RO) Result
pH value - 5-7.5 7.14 PASS
Hardness mg/L (ppm) ≤ 2 (Conversion from ≤ 0.02 mmol/L) 2 PASS
Conductivity (@ 25°C) µS/cm ≤ 5 51.6 FAIL
Chloride (Cl) mg/L (ppm) ≤ 0.2 15 FAIL
Iron mg/L (ppm) ≤ 0.1 0 PASS
Copper mg/L (ppm) ≤ 0.1 < 0.001 PASS
Silica (SiO₂) mg/L (ppm) ≤ 1 0 PASS
Phosphate (P₂O₅) mg/L (ppm) ≤ 0.5 0.05 PASS

Summary for Post-RO (DI/RO) Water:

The **Post-RO (DI/RO)** water **FAILS** the standard for two parameters:

The Hardness (2 mg/L) passes the converted standard (≤ 2 mg/L).



3. Almamlaka Labs — Aug 2024 Comparison (Your Images)

Note: RO Soft Water = Before RO (Pre‑RO / Softened),   RO Hot Water = After RO (Post‑RO).

Parameter Unit Before RO
(RO Soft Water)
After RO
(RO Hot Water)
pH (25°C)7.278.25
Total Dissolved Solids (TDS)mg/L (ppm)124.95.77
ConductivityµS/cm27613.18
Chloride (Cl)mg/L (ppm)604.5
Sulfate (SO₄)mg/L (ppm)11
Nitrate‑N (NO₃‑N)mg/L (ppm)0.20.5
Nitrite‑N (NO₂‑N)mg/L (ppm)23
Fluoride (F)mg/L (ppm)00
Total Phosphorus (PO₄)mg/L (ppm)0.0290.022
Calcium as CaCO₃mg/L (ppm)40
Calcium (Ca)mg/L (ppm)1.60
Magnesium as CaCO₃mg/L (ppm)40
Magnesium (Mg)mg/L (ppm)0.970
Iron (Fe, ferrous)mg/L (ppm)00.02
Cadmium (Cd)mg/L (ppm)<0.00005<0.00005
Total Hardnessmg/L (ppm)80
Total Chlorinemg/L (ppm)00.01
Silica (SiO₂)mg/L (ppm)0.32.9
OdorOdorlessOdorless

Quick Read


4. Multi-Report Comparative Trend Analysis (All Data)

The following tables consolidate all available water quality reports, sorted chronologically from August 2024 to October 2025, to show historical performance trends.


Table 1: Soft Water / Feed Water Trend Comparison

13/8/2024 (R.O SOFT) 13/8/2024 (TANK R.O SOFT) 24/9/2024 (Veolia Feed Water) 23/10/2025 (SOFT) Line Trend
7.27 7.18 8.02 6.68 Variable
124.9 124.7 164 200 Increasing (↑)
60 65 95 115 Increasing (↑)
1 1 N/R N/R Stable (≈)
8 8 0 2 Decreasing (↓)
0 0 0.18 0.06 Variable
276 277 355 433 Increasing (↑)

Table 2: Post-RO Product Water Trend Comparison

13/8/2024 (R.O HOT WATER) 13/8/2024 (TANK R.O WATER) 24/9/2024 (Veolia Hot Water) 23/10/2025 (R.O.01) Line Trend
8.25 8.4 8.13 7.14 Decreasing (↓)
5.77 5.4 5 23.3 Increasing (↑)
4.5 4.2 3 15 Increasing (↑)
1 1 N/R N/R Stable (≈)
0 0 0 2 Increasing (↑)
0.02 0.01 0.03 0 Variable
13.18 12.36 12 51.6 Increasing (↑)